About cookies on this site Our websites require some cookies to function properly (required). In addition, other cookies may be used with your consent to analyze site usage, improve the user experience and for advertising. For more information, please review your options. By visiting our website, you agree to our processing of information as described in IBM’sprivacy statement. To provide a smooth navigation, your cookie preferences will be shared across the IBM web domains listed here.
Publication
Microelectronic Engineering
Conference paper
Switchable cantilever fabrication for a novel time-of-flight scanning force microscope
Abstract
We describe a cantilever device for a novel time-of-flight scanning force microscopy (TOF-SFM) concept. The cantilever device consists of a switchable cantilever (SC), a microfabricated extraction electrode, and an Interlocking microstage. It allows quasi-simultaneous topographical and chemical imaging of a sample surface to be performed in the same way as with conventional scanning probe techniques. This is achieved by the micromachined SC with a bimorph actuator that provides a reasonable switching speed. Secondly, a short tip-electrode distance to minimize the ion extraction voltage can be realized by the help of the Interlock type assembling. The measured SC tip deflection is ∼ 100 μm at 35 mW, corresponding to an estimated heater temperature of ∼ 250°C. The maximum switching speed between the two modes is ∼ 50 ms, and the sensitivity ΔR/R of an integrated piezoresistive deflection sensor is ∼ 6.7 × 10-7/nm. The tip-electrode distance is only 10 mm. The TOF-SFM system is currently being integrated in an ultra-high-vacuum system to perform first experiments. © 2003 Elsevier Science B.V. All rights reserved.