Publication
Review of Scientific Instruments
Paper

Scanning superconducting quantum interference device susceptometry

View publication

Abstract

We report a scanning superconducting quantum interference device (SQUID) microsusceptometer with a spatial resolution of 8 μm, tested by measuring the susceptibility of individual 3 μm diam tin disks. Images of the disks agree well with numerical modeling based on the known geometry of the SQUID microsusceptometers. The low-field spin sensitivity between 1.5 and 6 K is 1 × 105μB/√Hz while scanning. © 2001 American Institute of Physics.