Aditya Kashyap, Julien F. Cors, et al.
Journal of Visualized Experiments
We present a new, general concept termed Hydrodynamic Thermal Confinement (HTC), and its implementation for the creation of microscale dynamic thermo-chemical microenvironments on biological surfaces. HTC is based on a scanning probe and operates under physiological conditions. The temperature can be regulated between 30° and 80 °C with ±0.2 °C precision and temperature ramps of 5 °C s-1 over a footprint of ∼50 μm × 80 μm in a volume of ∼50 × 80 × 15 μm3 (∼50 pl).
Aditya Kashyap, Julien F. Cors, et al.
Journal of Visualized Experiments
Aditya Kashyap, Julien F. Cors, et al.
MicroTAS 2016
Federico Paratore, Tal Zeidman Kalman, et al.
Analytical Chemistry
Xander F. van Kooten, Marianna Truman-Rosentsvit, et al.
Scientific Reports