Timothy C. Strand, Yigal Katzir
Applied Optics
An instrument capable of recording the amplitude and phase of reflected light with a phase resolution of better than X/3000 and the lateral resolution of a confocal scanning microscope was built. The instrument is based on a commercial microscope body and uses the regular interference contrast optics. The modifications consisted of adding a coherent (heterodyne) detector and a confocal laser scanning system. Two-dimensional surface images of amplitude, slope, and profile were taken with a step height resolution of typically 0.5-2 A. The instrument is described, and its characteristics for surface profilometry are discussed. © 1987 Optical Society of America.
Timothy C. Strand, Yigal Katzir
Applied Optics
Xu Wang, Jeffrey Mason, et al.
Journal of the Optical Society of America A: Optics and Image Science, and Vision
Ahmad Armand, Alexander A. Sawchuk, et al.
Applied Optics
Byron S. Lee, Timothy C. Strand
Applied Optics